Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communications to programmers and home monitoring units. Additionally, the Accent and Anthem pacemakers store the optional patient information without encryption. CVSS v3 base score: 3.1, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:N/A:N. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.
CVSS Information
N/A
Vulnerability Type
敏感数据加密缺失
Vulnerability Title
多款Abbott产品信息泄露漏洞
Vulnerability Description
Abbott Accent等都是美国雅培(Abbott)公司的心脏起搏器设备。 多款Abbott产品中存在信息泄露漏洞,该漏洞源于程序没有对病人信息进行加密。远程攻击者可利用该漏洞获取敏感信息。以下产品和版本受到影响:Abbott Accent F0B.0E.7E之前的版本;Accent MRI F10.08.6C之前的版本;Accent ST F10.08.6C之前的版本;Allure F14.07.80之前的版本;Anthem F0B.0E.7E之前的版本;Assurity F14.07.80之前的版