A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.
漏洞信息
N/A
漏洞
N/A
漏洞
Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump 安全漏洞
漏洞信息
Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump是英国史密斯医疗(Smiths Medical)公司的一款注射器输液泵。 Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump 1.1版本、1.5版本和1.6版本中存在安全漏洞,该漏洞源于程序使用了硬编码证书来自动创建无线网络连接。远程攻击者可利用该漏洞绕过安全限制。