Vulnerability Information
Although we use advanced large model technology, its output may still contain inaccurate or outdated information.Shenlong tries to ensure data accuracy, but please verify and judge based on the actual situation.
Vulnerability Title
N/A
Vulnerability Description
A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.
CVSS Information
N/A
Vulnerability Type
N/A
Vulnerability Title
Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump 安全漏洞
Vulnerability Description
Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump是英国史密斯医疗(Smiths Medical)公司的一款注射器输液泵。 Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump 1.1版本、1.5版本和1.6版本中存在安全漏洞,该漏洞源于程序使用了硬编码证书来自动创建无线网络连接。远程攻击者可利用该漏洞绕过安全限制。