Vulnerability Description
A Use of Hard-coded Password issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. Telnet on the pump uses hardcoded credentials, which can be used if the pump is configured to allow external communications. Smiths Medical assesses that it is not possible to upload files via Telnet and the impact of this vulnerability is limited to the communications module.
CVSS Information
N/A
Vulnerability Type
N/A
Vulnerability Title
Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump 安全漏洞
Vulnerability Description
Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump是英国史密斯医疗(Smiths Medical)公司的一款注射器输液泵。 Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump 1.1版本、1.5版本和1.6版本中存在安全绕过漏洞,该漏洞源于Telnet使用了硬编码证书。远程攻击者可利用该漏洞绕过安全限制。