Vulnerability Information
Although we use advanced large model technology, its output may still contain inaccurate or outdated information.Shenlong tries to ensure data accuracy, but please verify and judge based on the actual situation.
Vulnerability Title
N/A
Vulnerability Description
A vulnerability has been identified in RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems (All versions_without_ use of Siemens Healthineers Informatics products), RAPIDLab 1200 Series (All versions < V3.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions >= V3.0 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (V2.4.X_with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions =< V2.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 400 systems (All versions _with_ Siemens Healthineers Informatics products). A factory account with hardcoded password might allow attackers access to the device over port 5900/tcp. Successful exploitation requires no user interaction or privileges and impacts the confidentiality, integrity, and availability of the affected device. At the time of advisory publication, no public exploitation of this security vulnerability is known. Siemens Healthineers confirms the security vulnerability and provides mitigations to resolve the security issue.
CVSS Information
N/A
Vulnerability Type
使用硬编码的凭证
Vulnerability Title
多款Siemens产品安全漏洞
Vulnerability Description
Siemens RAPIDLab 1200 systems、RAPIDPoint 400 systems和RAPIDPoint 500 systems都是德国西门子(Siemens)公司的产品。Siemens RAPIDLab 1200 systems是一款具有血液、呼吸监控功能的重症监护解决方案。RAPIDPoint 400 systems和RAPIDPoint 500 systems都是不同系列的用于在临床中分析血夜中的电解质、葡萄糖、血细胞比容和新生儿胆红素的解决方案。 多款Siemens产品中存在
CVSS Information
N/A
Vulnerability Type
N/A