Vulnerability Title
Medtronic MyCareLink Smart Heap-based Buffer Overflow
Vulnerability Description
Medtronic MyCareLink Smart 25000 is vulnerable when an authenticated attacker runs a debug command, which can be sent to the patient reader and cause a heap overflow event within the MCL Smart Patient Reader software stack. The heap overflow could allow an attacker to remotely execute code on the MCL Smart Patient Reader, potentially leading to control of the device
CVSS Information
CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:C/C:H/I:H/A:H
Vulnerability Type
堆缓冲区溢出
Vulnerability Title
Medtronic MyCareLink Smart Model 25000 Patient Reader 安全漏洞
Vulnerability Description
Medtronic MyCareLink Smart Model 25000 Patient Reader是印度Medtronic公司的一个用于医疗行业可与医疗设备进行可视化交互的阅读器。 MCL Smart Patient Reader 存在安全漏洞,该漏洞源于当经过身份验证的攻击者运行调试命令时,受影响的产品就容易受到攻击,该命令可以发送给患者读取器,并导致MCL Smart患者读取器软件堆栈中的堆溢出事件。堆溢出可能允许攻击者远程执行MCL Smart Patient阅读器上的代码,从而可能导致对设
CVSS Information
N/A
Vulnerability Type
N/A