漏洞信息
尽管我们使用了先进的大模型技术,但其输出仍可能包含不准确或过时的信息。神龙努力确保数据的准确性,但请您根据实际情况进行核实和判断。
漏洞
N/A
漏洞信息
Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 contains a hard-coded password, which provides access to basic biomedical information, limited device settings, and network configuration of the WBM, if connected. The hard-coded password may allow an attacker with physical access to the device to access management functions to make unauthorized configuration changes to biomedical settings such as turn on and off wireless connections and the phase-complete audible alarm that indicates the end of an infusion phase. Baxter has released a new version of the SIGMA Spectrum Infusion System, version 8, which incorporates hardware and software changes.
漏洞信息
N/A
漏洞
使用硬编码的口令
漏洞
Baxter SIGMA Spectrum Infusion System 安全漏洞
漏洞信息
Baxter SIGMA Spectrum Infusion System是美国百特(Baxter)公司的一套智能输液系统。 Baxter SIGMA Spectrum Infusion System中存在本地安全绕过漏洞。本地攻击者可利用该漏洞绕过安全限制,执行未授权操作。