Vulnerability Information
Although we use advanced large model technology, its output may still contain inaccurate or outdated information.Shenlong tries to ensure data accuracy, but please verify and judge based on the actual situation.
Vulnerability Title
N/A
Vulnerability Description
Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
CVSS Information
N/A
Vulnerability Type
使用硬编码的口令
Vulnerability Title
多款Hospira产品安全漏洞
Vulnerability Description
Hospira Plum A+ Infusion System、Plum A+3 Infusion System和Symbiq Infusion System都是美国赫升瑞(Hospira)公司的智能输液系统。 多款Hospira产品中存在安全绕过漏洞。攻击者可利用该漏洞绕过安全限制,获取受影响设备的访问权限。以下产品及版本受到影响:Hospira Plum A+ Infusion System 13.4及之前版本,Plum A+3 Infusion System 13.6及之前版本,Symbiq Inf