CVE-2017-9657

Quick assessment

Affected
Philips IntelliVue MX40 Patient Worn Monitor
Exploitation
No confirmed in-the-wild exploitation; assess based on exposure
Recommended action
Check the vendor advisory and references for a fixed version. If immediate upgrade is impossible, restrict exposure and increase monitoring.

Philips IntelliVue MX40 Patient Worn Monitor是荷兰飞利浦(Philips)公司的一款医疗监护设备。 Philips IntelliVue MX40 Patient Worn Monitor B.06.18之前的版本(仅限WLAN)中存在安全漏洞。当设备运行在802.11 WLAN时,攻击者可利用该漏洞发布802.11 Wi-Fi管理命令,造成拒绝服务。

AI Predicted 6.5 Difficulty: Hard EPSS 0.77% · P56

Possible ATT&CK Techniques 1 AI

T1056 · Input Capture
Get alerts for future matching vulnerabilities Log in to subscribe

I. Basic Information for CVE-2017-9657

Vulnerability Information

Have questions about the vulnerability? See if Shenlong's analysis helps!
View Shenlong Deep Dive ↗

Although we use advanced large model technology, its output may still contain inaccurate or outdated information.Shenlong tries to ensure data accuracy, but please verify and judge based on the actual situation.

Vulnerability Title
N/A
Source: CVE Program / CVE List V5
Vulnerability Description
Under specific 802.11 network conditions, a partial re-association of the Philips IntelliVue MX40 Version B.06.18 WLAN monitor to the central monitoring station is possible. In this state, the central monitoring station can indicate the MX40 is not connected or associated to the central monitor, and thus should be operating in local monitoring mode (local audio-on, screen-on), but the MX40 WLAN itself can instead still be operating in telemetry mode (local audio-off, screen-off). If a patient experiences an alarm event and clinical staff expects the MX40 to provide local alarming when it is not available from the local device, a delay of treatment can occur. CVSS v3 base score: 6.5, CVSS vector string: AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H. Philips has released software update, Version B.06.18, to fix the improper cleanup on thrown exception vulnerability, and implement mitigations to reduce the risk associated with the improper handling of exceptional conditions vulnerability. The software update implements messaging and alarming on the MX40 and at the central monitoring station, when the MX40 disconnects from the access point.
Source: CVE Program / CVE List V5
CVSS Information
N/A
Source: CVE Program / CVE List V5
Vulnerability Type
抛出异常的清理不恰当
Source: CVE Program / CVE List V5
Vulnerability Title
Philips IntelliVue MX40 Patient Worn Monitor 安全漏洞
Source: CNNVD (China National Vulnerability Database)
Vulnerability Description
Philips IntelliVue MX40 Patient Worn Monitor是荷兰飞利浦(Philips)公司的一款医疗监护设备。 Philips IntelliVue MX40 Patient Worn Monitor B.06.18之前的版本(仅限WLAN)中存在安全漏洞。当设备运行在802.11 WLAN时,攻击者可利用该漏洞发布802.11 Wi-Fi管理命令,造成拒绝服务。
Source: CNNVD (China National Vulnerability Database)
CVSS Information
N/A
Source: CNNVD (China National Vulnerability Database)
Vulnerability Type
N/A
Source: CNNVD (China National Vulnerability Database)

Affected Products

Vendor Product Affected Versions CPE Subscribe
Philips IntelliVue MX40 Patient Worn Monitor IntelliVue MX40 Patient Worn Monitor (WLAN only), all versions prior to Version B.06.18 -