CVE-2018-14781— Medtronic MiniMed MMT-500/MMT-503 Remote Controllers Authentication Bypass by Capture-replay

Quick assessment

Affected
Medtronic MMT- 508 - MiniMed pump
Exploitation
No confirmed in-the-wild exploitation; assess based on exposure
Recommended action
Check the vendor advisory and references for a fixed version. If immediate upgrade is impossible, restrict exposure and increase monitoring.

Medtronic MMT-508 MiniMed insulin pump等都是美国Medtronic公司的不同型号的胰岛素泵。 多款Medtronic产品中存在授权问题漏洞。在配对远程控制器并启用‘easy bolus’和‘remote bolus’选项时,攻击者可利用该漏洞捕获控制器和pump之间传递的信息并进行胰岛素注射。以下产品受到影响:Medtronic MMT - 508 MiniMed insulin pump;MMT - 522 Paradigm REAL-TIME;MMT - 722

CVSS 5.3 · Medium EPSS 0.71% · P50
Get alerts for future matching vulnerabilities Log in to subscribe

I. Basic Information for CVE-2018-14781

Vulnerability Information

Have questions about the vulnerability? See if Shenlong's analysis helps!
View Shenlong Deep Dive ↗

Although we use advanced large model technology, its output may still contain inaccurate or outdated information.Shenlong tries to ensure data accuracy, but please verify and judge based on the actual situation.

Vulnerability Title
Medtronic MiniMed MMT-500/MMT-503 Remote Controllers Authentication Bypass by Capture-replay
Source: CVE Program / CVE List V5
Vulnerability Description
Medtronic MiniMed MMT devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.
Source: CVE Program / CVE List V5
CVSS Information
CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:N
Source: CVE Program / CVE List V5
Vulnerability Type
使用捕获-重放进行的认证绕过
Source: CVE Program / CVE List V5
Vulnerability Title
多款Medtronic产品授权问题漏洞
Source: CNNVD (China National Vulnerability Database)
Vulnerability Description
Medtronic MMT-508 MiniMed insulin pump等都是美国Medtronic公司的不同型号的胰岛素泵。 多款Medtronic产品中存在授权问题漏洞。在配对远程控制器并启用‘easy bolus’和‘remote bolus’选项时,攻击者可利用该漏洞捕获控制器和pump之间传递的信息并进行胰岛素注射。以下产品受到影响:Medtronic MMT - 508 MiniMed insulin pump;MMT - 522 Paradigm REAL-TIME;MMT - 722
Source: CNNVD (China National Vulnerability Database)
CVSS Information
N/A
Source: CNNVD (China National Vulnerability Database)
Vulnerability Type
N/A
Source: CNNVD (China National Vulnerability Database)

Affected Products

II. Public POCs for CVE-2018-14781

# POC Description Source Link Shenlong Link