I. Basic Information for CVE-2019-6538
Vulnerability Information

Have questions about the vulnerability? See if Shenlong's analysis helps!
View Shenlong Deep Dive ↗

Although we use advanced large model technology, its output may still contain inaccurate or outdated information.Shenlong tries to ensure data accuracy, but please verify and judge based on the actual situation.

Vulnerability Title
Medtronic Conexus Radio Frequency Telemetry Protocol Improper Access Control
Source: NVD (National Vulnerability Database)
Vulnerability Description
The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product’s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.
Source: NVD (National Vulnerability Database)
CVSS Information
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:C/C:N/I:H/A:H
Source: NVD (National Vulnerability Database)
Vulnerability Type
访问控制不恰当
Source: NVD (National Vulnerability Database)
Vulnerability Title
多款Medtronic产品访问控制错误漏洞
Source: CNNVD (China National Vulnerability Database)
Vulnerability Description
多款Medtronic产品中存在访问控制错误漏洞,该漏洞源于Conexus遥测协议没有执行授权或身份验证。攻击者可利用该漏洞注入,修改或拦截遥测通信中的输入。以下产品和版本受到影响:MyCareLink Monitor 24950版本,24952版本;CareLink Monitor 2490C版本;CareLink 2090 Programmer;Amplia CRT-D;Claria CRT-D;Compia CRT-D;Concerto CRT-D;Concerto II CRT-D;Consult
Source: CNNVD (China National Vulnerability Database)
CVSS Information
N/A
Source: CNNVD (China National Vulnerability Database)
Vulnerability Type
N/A
Source: CNNVD (China National Vulnerability Database)
Affected Products
VendorProductAffected VersionsCPESubscribe
MedtronicConexus Radio Frequency Telemetry Protocol All versions -
MedtronicMyCareLink Monitor 24950 -
MedtronicCareLink Monitor 2490C -
MedtronicCareLink 2090 Programmer All versions -
MedtronicAmplia CRT-D All versions -
MedtronicClaria CRT-D All versions -
MedtronicCompia CRT-D All versions -
MedtronicConcerto CRT-D All versions -
MedtronicConcerto II CRT-D All versions -
MedtronicConsulta CRT-D All versions -
MedtronicEvera ICD All versions -
MedtronicMaximo II CRT-D All versions -
MedtronicMaximo II ICD All versions -
MedtronicMirro ICD All versions -
MedtronicNayamed ND ICD All versions -
MedtronicPrimo ICD All versions -
MedtronicProtecta ICD, Protecta CRT-D All versions -
MedtronicSecura ICD All versions -
MedtronicVirtuoso ICD All versions -
MedtronicVirtuoso II ICD All versions -
MedtronicVisia AF ICD All versions -
MedtronicViva CRT-D All versions -
MedtronicBrava CRT-D All versions -
MedtronicMirro MRI ICD All versions -
II. Public POCs for CVE-2019-6538
#POC DescriptionSource LinkShenlong Link
AI-Generated POCPremium

No public POC found.

Login to generate AI POC
III. Intelligence Information for CVE-2019-6538
Please Login to view more intelligence information
IV. Related Vulnerabilities
V. Comments for CVE-2019-6538

No comments yet


Leave a comment