支持本站 — 捐款将帮助我们持续运营

目标: 1000 元,已筹: 1000

100.0%
获取后续新漏洞提醒登录后订阅
一、 漏洞 CVE-2019-6540 基础信息
漏洞信息

对漏洞内容有疑问?看看神龙的深度分析是否有帮助!
查看神龙十问 ↗

尽管我们使用了先进的大模型技术,但其输出仍可能包含不准确或过时的信息。神龙努力确保数据的准确性,但请您根据实际情况进行核实和判断。

漏洞
Medtronic Conexus Radio Frequency Telemetry Protocol Cleartext Transmission of Sensitive Information
来源: 美国国家漏洞数据库 NVD
漏洞信息
The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement encryption. An attacker with adjacent short-range access to a target product can listen to communications, including the transmission of sensitive data.
来源: 美国国家漏洞数据库 NVD
漏洞信息
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
来源: 美国国家漏洞数据库 NVD
漏洞
敏感数据的明文传输
来源: 美国国家漏洞数据库 NVD
漏洞
多款Medtronic产品安全漏洞
来源: 中国国家信息安全漏洞库 CNNVD
漏洞信息
多款Medtronic产品中存在安全漏洞,该漏洞源于Conexus遥测协议没有执行加密操作。攻击者可利用该漏洞监听通信(包括传输的敏感信息)。以下产品和版本受到影响:MyCareLink Monitor 24950版本,24952版本;CareLink Monitor 2490C版本;CareLink 2090 Programmer;Amplia CRT-D;Claria CRT-D;Compia CRT-D;Concerto CRT-D;Concerto II CRT-D;Consulta CRT-D;
来源: 中国国家信息安全漏洞库 CNNVD
漏洞信息
N/A
来源: 中国国家信息安全漏洞库 CNNVD
漏洞
N/A
来源: 中国国家信息安全漏洞库 CNNVD
受影响产品
厂商产品影响版本CPE订阅
MedtronicConexus Radio Frequency Telemetry Protocol All versions -
MedtronicMyCareLink Monitor 24950 -
MedtronicCareLink Monitor 2490C -
MedtronicCareLink 2090 Programmer All versions -
MedtronicAmplia CRT-D All versions -
MedtronicClaria CRT-D All versions -
MedtronicCompia CRT-D All versions -
MedtronicConcerto CRT-D All versions -
MedtronicConcerto II CRT-D All versions -
MedtronicConsulta CRT-D All versions -
MedtronicEvera ICD All versions -
MedtronicMaximo II CRT-D All versions -
MedtronicMaximo II ICD All versions -
MedtronicMirro ICD All versions -
MedtronicNayamed ND ICD All versions -
MedtronicPrimo ICD All versions -
MedtronicProtecta ICD, Protecta CRT-D All versions -
MedtronicSecura ICD All versions -
MedtronicVirtuoso ICD All versions -
MedtronicVirtuoso II ICD All versions -
MedtronicVisia AF ICD All versions -
MedtronicViva CRT-D All versions -
MedtronicBrava CRT-D All versions -
MedtronicMirro MRI ICD All versions -
二、漏洞 CVE-2019-6540 的公开POC
#POC 描述源链接神龙链接
AI 生成 POC高级

未找到公开 POC。

登录以生成 AI POC
三、漏洞 CVE-2019-6540 的情报信息
登录 三、漏洞 %(cve_id)s 的情报信息
新漏洞
四、漏洞 CVE-2019-6540 的评论

暂无评论


发表评论