Vulnerability Information
Although we use advanced large model technology, its output may still contain inaccurate or outdated information.Shenlong tries to ensure data accuracy, but please verify and judge based on the actual situation.
Vulnerability Title
Philips Patient Monitoring Devices Improper Handling of Length Parameter Inconsistency
Vulnerability Description
In Patient Information Center iX (PICiX) Versions C.02, C.03, the software parses a formatted message or structure but does not handle or incorrectly handles a length field that is inconsistent with the actual length of the associated data, causing the application on the surveillance station to restart.
CVSS Information
N/A
Vulnerability Type
长度参数不一致性处理不恰当
Vulnerability Title
Patient Information 安全漏洞
Vulnerability Description
Patient Information 中存在漏洞。该漏洞源于在电子表格软件打开文件时可能解释为命令的特殊元素。以下是受影响的产品及版本:B.02, C.02, C.03。
CVSS Information
N/A
Vulnerability Type
N/A