漏洞信息
尽管我们使用了先进的大模型技术,但其输出仍可能包含不准确或过时的信息。神龙努力确保数据的准确性,但请您根据实际情况进行核实和判断。
Vulnerability Title
N/A
Vulnerability Description
A vulnerability has been identified in DCA Vantage Analyzer (All versions < V4.5 are affected by CVE-2020-7590. In addition, serial numbers < 40000 running software V4.4.0 are also affected by CVE-2020-15797). Affected devices use a hard-coded password to protect the onboard database. This could allow an attacker to read and or modify the onboard database. Successful exploitation requires direct physical access to the device.
CVSS Information
N/A
Vulnerability Type
使用硬编码的口令
Vulnerability Title
Siemens DCA Vantage Analyzer 安全漏洞
Vulnerability Description
Siemens DCA Vantage Analyzer是德国西门子(Siemens)公司的一款糖化血红蛋白/尿微量白蛋白分析仪。 DCA Vantage Analyzer 4.5之前版本存在安全漏洞,该漏洞源于使用硬编码的密码来保护板载数据库。攻击者可利用该漏洞读取和修改机载数据库。
CVSS Information
N/A
Vulnerability Type
N/A