Vulnerability Information
Although we use advanced large model technology, its output may still contain inaccurate or outdated information.Shenlong tries to ensure data accuracy, but please verify and judge based on the actual situation.
Vulnerability Title
Improper Access Control for Boston Scientific Zoom Latitude
Vulnerability Description
A skilled attacker with physical access to the affected device can gain access to the hard disk drive of the device to change the telemetry region and could use this setting to interrogate or program an implantable device in any region in the world.
CVSS Information
CVSS:3.1/AV:P/AC:L/PR:N/UI:R/S:C/C:L/I:H/A:L
Vulnerability Type
访问控制不恰当
Vulnerability Title
Boston Scientific Zoom Latitude Programmer/Recorder/Monitor Model 3120 访问控制错误漏洞
Vulnerability Description
Boston Scientific Zoom Latitude Programmer/Recorder/Monitor Model 3120(Zoom Latitude Prm Model 3120)是美国Boston Scientific公司的一种便携式心律管理 (Crm) 编程系统。用于询问、监测和编程Boston Scientific公司的植入式脉冲发生器 (Ipg)。 Boston Scientific Zoom Latitude Programmer/Recorder/Monitor Model
CVSS Information
N/A
Vulnerability Type
N/A