漏洞信息
尽管我们使用了先进的大模型技术,但其输出仍可能包含不准确或过时的信息。神龙努力确保数据的准确性,但请您根据实际情况进行核实和判断。
Vulnerability Title
Dräger Infinity M540 VG4.1.1 Spoofed Network Message Handling DoS/Tampering
Vulnerability Description
Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors versions VG4.1.1, VG4.0.3, and lower (with VG4.2 partially affected) contain a network message handling vulnerability that allows remote attackers to inject spoofed or tampered data and cause denial-of-service conditions. Attackers can compromise network communications to modify device settings such as alarm states or alarm limits, or overwhelm the system with excessive network traffic causing the Cockpit or M540 to reboot and lose network functionality.
CVSS Information
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H
Vulnerability Type
对数据真实性的验证不充分
Vulnerability Title
Dräger Infinity Acute Care System和Dräger Standalone Infinity M540 patient monitor 数据伪造问题漏洞
Vulnerability Description
Dräger Infinity Acute Care System和Dräger Standalone Infinity M540 patient monitor都是德国Dräger公司的产品。Dräger Infinity Acute Care System是一个集患者监护、临床工作站和医疗信息集成于一体的急症监护平台。Dräger Standalone Infinity M540 patient monitor是一款模块化床旁患者监护仪。 Dräger Infinity Acute Care Syst
CVSS Information
N/A
Vulnerability Type
N/A