Vulnerability Title
Medtronic Micro Clinician & InterStim X Clinician App Password Reset Issue
来源: 美国国家漏洞数据库 NVD
Vulnerability Description
Medtronic identified that the Pelvic Health clinician apps, which are installed on the Smart Programmer mobile device, have a password vulnerability that requires a security update to fix. Not updating could potentially result in unauthorized control of the clinician therapy application, which has greater control over therapy parameters than the patient app. Changes still cannot be made outside of the established therapy parameters of the programmer. For unauthorized access to occur, an individual would need physical access to the Smart Programmer.
来源: 美国国家漏洞数据库 NVD
CVSS Information
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:H/A:H
来源: 美国国家漏洞数据库 NVD
Vulnerability Type
未经验证的口令修改
来源: 美国国家漏洞数据库 NVD
Vulnerability Title
Medtronic Pelvic Health 授权问题漏洞
来源: 中国国家信息安全漏洞库 CNNVD
Vulnerability Description
Medtronic Pelvic Health是美国美敦力(Medtronic)公司的一个临床医生应用程序。 Medtronic Pelvic Health clinician 存在安全漏洞,该漏洞源于应用程序存在密码漏洞,攻击者利用该漏洞可以对该应用程序进行未经授权控制。
来源: 中国国家信息安全漏洞库 CNNVD
CVSS Information
N/A
来源: 中国国家信息安全漏洞库 CNNVD
Vulnerability Type
N/A
来源: 中国国家信息安全漏洞库 CNNVD
受影响产品
厂商产品影响版本CPE订阅
MedtronicInsterStim Applications Micro Clinician ~ A51200 -