Goal Reached Thanks to every supporter — we hit 100%!

Goal: 1000 CNY · Raised: 1336 CNY

100%

CVE-2023-30561— Lack of Cryptographic Security of IUI Bus

Quick assessment

Affected
Becton Dickinson & Co BD Alarisâ„¢ Point-of-Care Unit (PCU) Model 8015
Exploitation
No confirmed in-the-wild exploitation; assess based on exposure
Recommended action
Check the vendor advisory and references for a fixed version. If immediate upgrade is impossible, restrict exposure and increase monitoring.

BD Alaris System with Guardrails Suite MX是美国碧迪医疗(BD)公司的一套医疗设备。 BD Alaris System with Guardrails Suite MX 存在安全漏洞,该漏洞源于 PCU 及其模块之间的数据传输不安全。具有物理访问权限的威胁参与者可能会在输注运行时通过连接特制设备来读取或修改数据。

CVSS 6.1 · Medium EPSS 0.19% · P9

Possible ATT&CK Techniques 1 AI

T1217 · Browser Information Discovery
Get alerts for future matching vulnerabilities Log in to subscribe

I. Basic Information for CVE-2023-30561

Vulnerability Information

Have questions about the vulnerability? See if Shenlong's analysis helps!
View Shenlong Deep Dive ↗

Although we use advanced large model technology, its output may still contain inaccurate or outdated information.Shenlong tries to ensure data accuracy, but please verify and judge based on the actual situation.

Vulnerability Title
Lack of Cryptographic Security of IUI Bus
Source: CVE Program / CVE List V5
Vulnerability Description
The data flowing between the PCU and its modules is insecure. A threat actor with physical access could potentially read or modify data by attaching a specially crafted device while an infusion is running.
Source: CVE Program / CVE List V5
CVSS Information
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:H
Source: CVE Program / CVE List V5
Vulnerability Type
敏感数据加密缺失
Source: CVE Program / CVE List V5
Vulnerability Title
BD Alaris System with Guardrails Suite MX 安全漏洞
Source: CNNVD (China National Vulnerability Database)
Vulnerability Description
BD Alaris System with Guardrails Suite MX是美国碧迪医疗(BD)公司的一套医疗设备。 BD Alaris System with Guardrails Suite MX 存在安全漏洞,该漏洞源于 PCU 及其模块之间的数据传输不安全。具有物理访问权限的威胁参与者可能会在输注运行时通过连接特制设备来读取或修改数据。
Source: CNNVD (China National Vulnerability Database)
CVSS Information
N/A
Source: CNNVD (China National Vulnerability Database)
Vulnerability Type
N/A
Source: CNNVD (China National Vulnerability Database)

Affected Products

Vendor Product Affected Versions CPE Subscribe
Becton Dickinson & Co BD Alarisâ„¢ Point-of-Care Unit (PCU) Model 8015 0 ~ 12.1.3 -

II. Public POCs for CVE-2023-30561

# POC Description Source Link Shenlong Link
AI-Generated POC Premium

No public POC found.

Login to generate AI POC

III. Intelligence Information for CVE-2023-30561

登录查看更多情报信息。

Other References for CVE-2023-30561 (1)

Same Patch Batch · Becton Dickinson & Co · 2023-07-13 · 5 CVEs total

CVE-2023-30563 8.2 HIGH Stored Cross-Site Scripting on User Import Functionality
CVE-2023-30564 6.9 MEDIUM Stored Cross-Site Scripting on Device Import Functionality
CVE-2023-30565 3.5 LOW CQI Data Sniffing
CVE-2023-30562 3.0 LOW Lack of Dataset Integrity Checking

IV. Related Vulnerabilities

V. Comments for CVE-2023-30561

No comments yet


Leave a comment