Goal Reached Thanks to every supporter — we hit 100%!

Goal: 1000 CNY · Raised: 1359 CNY

100%

CVE-2024-1630— Path traversal vulnerability in “getAllFolderContents” function of Common Service Desktop, a GE HealthCare ultrasound device component

Quick assessment

Affected
GE HealthCare Venue
Exploitation
No confirmed in-the-wild exploitation; assess based on exposure
Recommended action
Check the vendor advisory and references for a fixed version. If immediate upgrade is impossible, restrict exposure and increase monitoring.

GE General Electric Healthcare Ultrasound是美国通用电气(GE)公司的一个医学声波检测设备。 General Electric Healthcare Ultrasound存在路径遍历漏洞。攻击者利用该漏洞可以访问存储在web根文件夹之外的文件和目录。

CVSS 7.7 · High EPSS 0.28% · P19

Possible ATT&CK Techniques 1 AI

T1083 · File and Directory Discovery
Get alerts for future matching vulnerabilities Log in to subscribe

I. Basic Information for CVE-2024-1630

Vulnerability Information

Have questions about the vulnerability? See if Shenlong's analysis helps!
View Shenlong Deep Dive ↗

Although we use advanced large model technology, its output may still contain inaccurate or outdated information.Shenlong tries to ensure data accuracy, but please verify and judge based on the actual situation.

Vulnerability Title
Path traversal vulnerability in “getAllFolderContents” function of Common Service Desktop, a GE HealthCare ultrasound device component
Source: CVE Program / CVE List V5
Vulnerability Description
Path traversal vulnerability in “getAllFolderContents” function of Common Service Desktop, a GE HealthCare ultrasound device component
Source: CVE Program / CVE List V5
CVSS Information
CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:N
Source: CVE Program / CVE List V5
Vulnerability Type
对路径名的限制不恰当(路径遍历)
Source: CVE Program / CVE List V5
Vulnerability Title
General Electric Healthcare Ultrasound 路径遍历漏洞
Source: CNNVD (China National Vulnerability Database)
Vulnerability Description
GE General Electric Healthcare Ultrasound是美国通用电气(GE)公司的一个医学声波检测设备。 General Electric Healthcare Ultrasound存在路径遍历漏洞。攻击者利用该漏洞可以访问存储在web根文件夹之外的文件和目录。
Source: CNNVD (China National Vulnerability Database)
CVSS Information
N/A
Source: CNNVD (China National Vulnerability Database)
Vulnerability Type
N/A
Source: CNNVD (China National Vulnerability Database)

Affected Products

Vendor Product Affected Versions CPE Subscribe
GE HealthCare Venue R1 -
GE HealthCare Venue Go R2 -
GE HealthCare Venue Fit R3 ~ R3.3 -
GE HealthCare LOGIQ e R7 ~ R9.1.4 -
GE HealthCare LOGIQ He 0 ~ R9.3.1 -
GE HealthCare Vivid E E95 ~ 206 -
GE HealthCare Vivid T T8 ~ 206 -
GE HealthCare Vivid iq 0 ~ 206 -
GE HealthCare Voluson Expert 16 0 -
GE HealthCare Voluson Expert 18 0 -
GE HealthCare Voluson Expert 22 0 -
GE HealthCare Voluson SWIFT 0 -
GE HealthCare LOGIQ E10 0 ~ R3.2.0 -
GE HealthCare LOGIQ E10s 0 ~ R3.2.0 -
GE HealthCare LOGIQ Fortis 0 ~ R3.2.0 -

II. Public POCs for CVE-2024-1630

# POC Description Source Link Shenlong Link
AI-Generated POC Premium

No public POC found.

Login to generate AI POC

III. Intelligence Information for CVE-2024-1630

请登录查看更多情报信息。

Other References for CVE-2024-1630 (1)

Same Patch Batch · GE HealthCare · 2024-05-14 · 9 CVEs total

CVE-2024-27107 9.6 CRITICAL Weak account password in GE HealthCare EchoPAC products
CVE-2024-1628 8.4 HIGH OS command injection vulnerabilities in GE HealthCare ultrasound devices
CVE-2024-27110 8.4 HIGH Elevation of privilege vulnerability in GE HealthCare EchoPAC products
CVE-2024-27109 7.6 HIGH Insufficiently protected credentials in GE HealthCare EchoPAC products
CVE-2024-1486 7.4 HIGH Elevation of privileges via misconfigured access control list in GE HealthCare ultrasound
CVE-2024-27108 6.8 MEDIUM Non privileged access to critical file vulnerability in GE HealthCare EchoPAC products
CVE-2024-1629 6.2 MEDIUM Path traversal vulnerability in “deleteFiles” function of Common Service Desktop, a GE Hea
CVE-2024-27106 5.7 MEDIUM Vulnerable data in transit in GE HealthCare EchoPAC products

IV. Related Vulnerabilities

V. Comments for CVE-2024-1630

No comments yet


Leave a comment