CVE-2024-48974— Life2000 Ventilator does not perform proper file integrity checks when adopting firmware updates

Quick assessment

Affected
Baxter Life2000 Ventilation System
Exploitation
No confirmed in-the-wild exploitation; assess based on exposure
Recommended action
Check the vendor advisory and references for a fixed version. If immediate upgrade is impossible, restrict exposure and increase monitoring.

Baxter Life2000是Baxter公司的一款无需面罩的无创呼吸机。 Baxter Life2000 06.08.00.00及之前版本存在安全漏洞,该漏洞源于在采用固件更新时文件完整性检查不当,使得攻击者可以推送受损或非法的固件文件,可能会破坏设备的功能或导致未经授权的信息泄露。

CVSS 9.3 · Critical EPSS 0.15% · P4

Possible ATT&CK Techniques 1 AI

T1195 · Supply Chain Compromise
Get alerts for future matching vulnerabilities Log in to subscribe

I. Basic Information for CVE-2024-48974

Vulnerability Information

Have questions about the vulnerability? See if Shenlong's analysis helps!
View Shenlong Deep Dive ↗

Although we use advanced large model technology, its output may still contain inaccurate or outdated information.Shenlong tries to ensure data accuracy, but please verify and judge based on the actual situation.

Vulnerability Title
Life2000 Ventilator does not perform proper file integrity checks when adopting firmware updates
Source: CVE Program / CVE List V5
Vulnerability Description
The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This could disrupt the function of the device and/or cause unauthorized information disclosure.
Source: CVE Program / CVE List V5
CVSS Information
CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
Source: CVE Program / CVE List V5
Vulnerability Type
下载代码缺少完整性检查
Source: CVE Program / CVE List V5
Vulnerability Title
Baxter Life2000 安全漏洞
Source: CNNVD (China National Vulnerability Database)
Vulnerability Description
Baxter Life2000是Baxter公司的一款无需面罩的无创呼吸机。 Baxter Life2000 06.08.00.00及之前版本存在安全漏洞,该漏洞源于在采用固件更新时文件完整性检查不当,使得攻击者可以推送受损或非法的固件文件,可能会破坏设备的功能或导致未经授权的信息泄露。
Source: CNNVD (China National Vulnerability Database)
CVSS Information
N/A
Source: CNNVD (China National Vulnerability Database)
Vulnerability Type
N/A
Source: CNNVD (China National Vulnerability Database)

Shenlong Deep Dive — AI Deep Analysis

10-question deep dive: root cause, exploitation, mitigation, urgency. Read summary free, full version requires login.

Affected Products

Vendor Product Affected Versions CPE Subscribe
Baxter Life2000 Ventilation System 06.08.00.00 and prior -

II. Public POCs for CVE-2024-48974

# POC Description Source Link Shenlong Link
AI-Generated POC Premium

No public POC found.

Login to generate AI POC

III. Intelligence Information for CVE-2024-48974

请登录查看更多情报信息。

Vendor Advisories for CVE-2024-48974 (1)

Same Patch Batch · Baxter · 2024-11-14 · 8 CVEs total

CVE-2024-48966 10.0 CRITICAL Life2000 service tools for test and calibration do not support user authentication
CVE-2024-48967 10.0 CRITICAL Life2000 ventilator and Service PC lack sufficient audit logging capabilities
CVE-2024-9832 9.3 CRITICAL No limit on failed login attempts with Clinician Password or Serial Number Clinician Passw
CVE-2024-9834 9.3 CRITICAL Improper data protection on Life2000 ventilator serial interface
CVE-2024-48971 9.3 CRITICAL Clinician Password and Serial Number Clinician Password are hard-coded in Life2000 Ventila
CVE-2024-48970 9.3 CRITICAL Life2000 Ventilator microcontroller lacks memory protection
CVE-2024-48973 9.3 CRITICAL Debug port on Life2000 Ventilator serial interface is enabled by default