Vulnerability Information
Although we use advanced large model technology, its output may still contain inaccurate or outdated information.Shenlong tries to ensure data accuracy, but please verify and judge based on the actual situation.
Vulnerability Title
N/A
Vulnerability Description
The "update" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function triggers if the 'C' button is pressed at a specific time during the boot process. If an attacker is able to control or impersonate this IP address, they could upload and overwrite files on the device.
CVSS Information
N/A
Vulnerability Type
隐藏功能
Vulnerability Title
Contec Health CMS8000 Patient Monitor 安全漏洞
Vulnerability Description
Contec Health CMS8000 Patient Monitor是日本Contec公司的一款生命体征病人监护仪。 Contec Health CMS8000 Patient Monitor存在安全漏洞,该漏洞源于更新二进制文件会尝试安装到硬编码的可路由IP地址,从而绕过现有的设备网络设置。
CVSS Information
N/A
Vulnerability Type
N/A