漏洞信息
尽管我们使用了先进的大模型技术,但其输出仍可能包含不准确或过时的信息。神龙努力确保数据的准确性,但请您根据实际情况进行核实和判断。
Vulnerability Title
Eppendorf BioFlo 320 Use of hard-coded password
Vulnerability Description
Eppendorf BioFlo 320 is vulnerable to due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted.
CVSS Information
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Vulnerability Type
使用硬编码的口令
Vulnerability Title
Eppendorf BioFlo 320 安全漏洞
Vulnerability Description
Eppendorf BioFlo 320是德国Eppendorf公司的一款实验室生物反应器控制系统。 Eppendorf BioFlo 320存在安全漏洞,该漏洞源于VNC服务器使用硬编码密码,可能导致远程攻击者通过已知网络地址获得用户界面完全控制权,VNC流量未加密。
CVSS Information
N/A
Vulnerability Type
N/A