Vulnerability Information
Although we use advanced large model technology, its output may still contain inaccurate or outdated information.Shenlong tries to ensure data accuracy, but please verify and judge based on the actual situation.
Vulnerability Title
N/A
Vulnerability Description
Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 is remotely accessible via Port 22/SSH without authentication. A remote attacker may be able to make unauthorized configuration changes to the WBM, as well as issue commands to access account credentials and shared keys. Baxter asserts that this vulnerability only allows access to features and functionality on the WBM and that the SIGMA Spectrum infusion pump cannot be controlled from the WBM. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.
CVSS Information
N/A
Vulnerability Type
CWE-592
Vulnerability Title
Baxter Wireless Battery Module 安全漏洞
Vulnerability Description
Baxter SIGMA Spectrum Infusion System是美国百特(Baxter)公司的一套智能输液系统。Baxter Wireless Battery(WBM)是其中的一个无线供电模块。 Baxter WBM中存在安全漏洞。攻击者可利用该漏洞获取未授权的访问权限和敏感信息,或绕过身份验证机制,获取受影响设备的访问权限。
CVSS Information
N/A
Vulnerability Type
N/A