CVE-2019-25719— Dräger Infinity M540 VG4.1.1 Spoofing and DoS via Network Message Handling

CVSS 8.6 · High

Affected Version Matrix 7

VendorProductVersion RangeStatus
DrägerInfinity Acute Care SystemVG4.1.1affected
VG4.0.3affected
lower than VG4.0.3affected
VG4.2unaffected
DrägerStandalone Infinity M540 patient monitorVG4.1.1affected
lower than VG4.1.1affected
VG4.2unaffected
Get alerts for future matching vulnerabilitiesLog in to subscribe

I. Basic Information for CVE-2019-25719

Vulnerability Information

Have questions about the vulnerability? See if Shenlong's analysis helps!
View Shenlong Deep Dive ↗

Although we use advanced large model technology, its output may still contain inaccurate or outdated information.Shenlong tries to ensure data accuracy, but please verify and judge based on the actual situation.

Vulnerability Title
Dräger Infinity M540 VG4.1.1 Spoofing and DoS via Network Message Handling
Source: NVD (National Vulnerability Database)
Vulnerability Description
Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.
Source: NVD (National Vulnerability Database)
CVSS Information
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H
Source: NVD (National Vulnerability Database)
Vulnerability Type
通信信道中传输过程中消息完整性的不正确执行
Source: NVD (National Vulnerability Database)

Affected Products

VendorProductAffected VersionsCPESubscribe
DrägerInfinity Acute Care System VG4.1.1 -
DrägerStandalone Infinity M540 patient monitor VG4.1.1 -

II. Public POCs for CVE-2019-25719

#POC DescriptionSource LinkShenlong Link
AI-Generated POCPremium

No public POC found.

Login to generate AI POC

III. Intelligence Information for CVE-2019-25719

登录查看更多情报信息。

Vendor Advisories for CVE-2019-25719 (1)

Same Patch Batch · Dräger · 2026-06-02 · 8 CVEs total

CVE-2021-44788.2 HIGHDräger CC-Vision Basic and CC-Vision E-Cal Out-of-Bounds Write via Malicious GDT File
CVE-2019-257227.6 HIGHDräger SC Monitoring Devices Hard-coded Credentials and DoS
CVE-2019-257246.5 MEDIUMDräger Infinity M300 VG2.x Network-Based Denial of Service
CVE-2019-257216.5 MEDIUMDräger Infinity M300 VG2.3.1 Network-Based Denial of Service
CVE-2019-257174.3 MEDIUMDräger Infinity Delta/Kappa Patient Monitors Unauthenticated Log File Disclosure
CVE-2019-257234.0 MEDIUMDräger Perseus A500 2.00-2.02 DoS via Medibus Interface
CVE-2021-44794.0 MEDIUMDräger Atlan A350 1.00-1.01 DoS via Medibus Interface