CVE-2019-6538— Medtronic Conexus Radio Frequency Telemetry Protocol Improper Access Control

Quick assessment

Affected
Medtronic Conexus Radio Frequency Telemetry Protocol
Exploitation
No confirmed in-the-wild exploitation; assess based on exposure
Recommended action
Check the vendor advisory and references for a fixed version. If immediate upgrade is impossible, restrict exposure and increase monitoring.

多款Medtronic产品中存在访问控制错误漏洞,该漏洞源于Conexus遥测协议没有执行授权或身份验证。攻击者可利用该漏洞注入,修改或拦截遥测通信中的输入。以下产品和版本受到影响:MyCareLink Monitor 24950版本,24952版本;CareLink Monitor 2490C版本;CareLink 2090 Programmer;Amplia CRT-D;Claria CRT-D;Compia CRT-D;Concerto CRT-D;Concerto II CRT-D;Consult

CVSS 9.3 · Critical EPSS 0.90% · P58
Get alerts for future matching vulnerabilities Log in to subscribe

I. Basic Information for CVE-2019-6538

Vulnerability Information

Have questions about the vulnerability? See if Shenlong's analysis helps!
View Shenlong Deep Dive ↗

Although we use advanced large model technology, its output may still contain inaccurate or outdated information.Shenlong tries to ensure data accuracy, but please verify and judge based on the actual situation.

Vulnerability Title
Medtronic Conexus Radio Frequency Telemetry Protocol Improper Access Control
Source: CVE Program / CVE List V5
Vulnerability Description
The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product’s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.
Source: CVE Program / CVE List V5
CVSS Information
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:C/C:N/I:H/A:H
Source: CVE Program / CVE List V5
Vulnerability Type
访问控制不恰当
Source: CVE Program / CVE List V5
Vulnerability Title
多款Medtronic产品访问控制错误漏洞
Source: CNNVD (China National Vulnerability Database)
Vulnerability Description
多款Medtronic产品中存在访问控制错误漏洞,该漏洞源于Conexus遥测协议没有执行授权或身份验证。攻击者可利用该漏洞注入,修改或拦截遥测通信中的输入。以下产品和版本受到影响:MyCareLink Monitor 24950版本,24952版本;CareLink Monitor 2490C版本;CareLink 2090 Programmer;Amplia CRT-D;Claria CRT-D;Compia CRT-D;Concerto CRT-D;Concerto II CRT-D;Consult
Source: CNNVD (China National Vulnerability Database)
CVSS Information
N/A
Source: CNNVD (China National Vulnerability Database)
Vulnerability Type
N/A
Source: CNNVD (China National Vulnerability Database)

Affected Products

Vendor Product Affected Versions CPE Subscribe
Medtronic Conexus Radio Frequency Telemetry Protocol All versions -
Medtronic MyCareLink Monitor 24950 -
Medtronic CareLink Monitor 2490C -
Medtronic CareLink 2090 Programmer All versions -
Medtronic Amplia CRT-D All versions -
Medtronic Claria CRT-D All versions -
Medtronic Compia CRT-D All versions -
Medtronic Concerto CRT-D All versions -
Medtronic Concerto II CRT-D All versions -
Medtronic Consulta CRT-D All versions -
Medtronic Evera ICD All versions -
Medtronic Maximo II CRT-D All versions -
Medtronic Maximo II ICD All versions -
Medtronic Mirro ICD All versions -
Medtronic Nayamed ND ICD All versions -
Medtronic Primo ICD All versions -
Medtronic Protecta ICD, Protecta CRT-D All versions -
Medtronic Secura ICD All versions -
Medtronic Virtuoso ICD All versions -
Medtronic Virtuoso II ICD All versions -
Medtronic Visia AF ICD All versions -
Medtronic Viva CRT-D All versions -
Medtronic Brava CRT-D All versions -
Medtronic Mirro MRI ICD All versions -

II. Public POCs for CVE-2019-6538

# POC Description Source Link Shenlong Link
AI-Generated POC Premium

No public POC found.

Login to generate AI POC

III. Intelligence Information for CVE-2019-6538

请登录查看更多情报信息。

Vendor Advisories for CVE-2019-6538 (1)

Other References for CVE-2019-6538 (1)

IV. Related Vulnerabilities

V. Comments for CVE-2019-6538

No comments yet


Leave a comment