CVE-2025-0683— Exposure of Private Personal Information to an Unauthorized Actor vulnerability in Contec Health CMS8000 Patient Monitor

Quick assessment

Affected
Contec Health CMS8000 Patient Monitor
Exploitation
No confirmed in-the-wild exploitation; assess based on exposure
Recommended action
Check the vendor advisory and references for a fixed version. If immediate upgrade is impossible, restrict exposure and increase monitoring.

Contec Health CMS8000 Patient Monitor是日本Contec公司的一款生命体征病人监护仪。 Contec Health CMS8000 Patient Monitor存在安全漏洞,该漏洞源于允许将纯文本患者数据传输到硬编码的公共 IP 地址。

CVSS 5.9 · Medium EPSS 0.79% · P55

Possible ATT&CK Techniques 1 AI

T1572 · Protocol Tunneling
Get alerts for future matching vulnerabilities Log in to subscribe

I. Basic Information for CVE-2025-0683

Vulnerability Information

Have questions about the vulnerability? See if Shenlong's analysis helps!
View Shenlong Deep Dive ↗

Although we use advanced large model technology, its output may still contain inaccurate or outdated information.Shenlong tries to ensure data accuracy, but please verify and judge based on the actual situation.

Vulnerability Title
Exposure of Private Personal Information to an Unauthorized Actor vulnerability in Contec Health CMS8000 Patient Monitor
Source: CVE Program / CVE List V5
Vulnerability Description
In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text patient data to a hard-coded public IP address when a patient is hooked up to the monitor. This could lead to a leakage of confidential patient data to any device with that IP address or an attacker in a machine-in-the-middle scenario.
Source: CVE Program / CVE List V5
CVSS Information
CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N
Source: CVE Program / CVE List V5
Vulnerability Type
侵犯隐私
Source: CVE Program / CVE List V5
Vulnerability Title
Contec Health CMS8000 Patient Monitor 安全漏洞
Source: CNNVD (China National Vulnerability Database)
Vulnerability Description
Contec Health CMS8000 Patient Monitor是日本Contec公司的一款生命体征病人监护仪。 Contec Health CMS8000 Patient Monitor存在安全漏洞,该漏洞源于允许将纯文本患者数据传输到硬编码的公共 IP 地址。
Source: CNNVD (China National Vulnerability Database)
CVSS Information
N/A
Source: CNNVD (China National Vulnerability Database)
Vulnerability Type
N/A
Source: CNNVD (China National Vulnerability Database)

Affected Products

Vendor Product Affected Versions CPE Subscribe
Contec Health CMS8000 Patient Monitor All versions -

II. Public POCs for CVE-2025-0683

# POC Description Source Link Shenlong Link
AI-Generated POC Premium

No public POC found.

Login to generate AI POC

III. Intelligence Information for CVE-2025-0683

请登录查看更多情报信息。

Vendor Advisories for CVE-2025-0683 (2)

Same Patch Batch · Contec Health · 2025-01-30 · 3 CVEs total

CVE-2024-12248 9.8 CRITICAL Out-of-bounds Write vulnerability in Contec Health CMS8000 Patient Monitor
CVE-2025-0626 7.5 HIGH Hidden Functionality vulnerability in Contec Health CMS8000 Patient Monitor