Vulnerability Information
Although we use advanced large model technology, its output may still contain inaccurate or outdated information.Shenlong tries to ensure data accuracy, but please verify and judge based on the actual situation.
Vulnerability Title
Credential Disclosure vulnerability in Radiometer Products
Vulnerability Description
A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system. Other related CVE's are CVE-2025-14095 & CVE-2025-14097. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Attacker requires physical access to the analyzer. Temporary work Around: Only authorized people can physically access the analyzer. Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.
CVSS Information
CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Vulnerability Type
使用硬编码的凭证
Vulnerability Title
Radiometer Products 安全漏洞
Vulnerability Description
Radiometer Products是丹麦Radiometer公司的一系列医疗诊断仪器。 Radiometer Products存在安全漏洞,该漏洞源于操作系统设计和凭据保护不足,可能导致物理访问者提取凭据信息。
CVSS Information
N/A
Vulnerability Type
N/A