Goal Reached Thanks to every supporter — we hit 100%!

Goal: 1000 CNY · Raised: 1359 CNY

100%

CVE-2025-4394— Medtronic MyCareLink Patient Monitor Unencrypted Filesystem Vulnerability

Quick assessment

Affected
Medtronic MyCareLink Patient Monitor 24950
Exploitation
No confirmed in-the-wild exploitation; assess based on exposure
Recommended action
Check the vendor advisory and references for a fixed version. If immediate upgrade is impossible, restrict exposure and increase monitoring.

Medtronic MyCareLink Patient Monitor是美国Medtronic开源的一款用于远程监测患者的监控系统。 Medtronic MyCareLink Patient Monitor 2025.6.25之前版本存在安全漏洞,该漏洞源于内部存储使用未加密文件系统,可能导致物理访问攻击者读取和修改文件。

CVSS 6.8 · Medium EPSS 0.19% · P8

Possible ATT&CK Techniques 1 AI

T1552.004 · Private Keys
Get alerts for future matching vulnerabilities Log in to subscribe

I. Basic Information for CVE-2025-4394

Vulnerability Information

Have questions about the vulnerability? See if Shenlong's analysis helps!
View Shenlong Deep Dive ↗

Although we use advanced large model technology, its output may still contain inaccurate or outdated information.Shenlong tries to ensure data accuracy, but please verify and judge based on the actual situation.

Vulnerability Title
Medtronic MyCareLink Patient Monitor Unencrypted Filesystem Vulnerability
Source: CVE Program / CVE List V5
Vulnerability Description
Medtronic MyCareLink Patient Monitor uses an unencrypted filesystem on internal storage, which allows an attacker with physical access to read and modify files. This issue affects MyCareLink Patient Monitor models 24950 and 24952: before June 25, 2025
Source: CVE Program / CVE List V5
CVSS Information
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Source: CVE Program / CVE List V5
Vulnerability Type
敏感数据的明文存储
Source: CVE Program / CVE List V5
Vulnerability Title
Medtronic MyCareLink Patient Monitor 安全漏洞
Source: CNNVD (China National Vulnerability Database)
Vulnerability Description
Medtronic MyCareLink Patient Monitor是美国Medtronic开源的一款用于远程监测患者的监控系统。 Medtronic MyCareLink Patient Monitor 2025.6.25之前版本存在安全漏洞,该漏洞源于内部存储使用未加密文件系统,可能导致物理访问攻击者读取和修改文件。
Source: CNNVD (China National Vulnerability Database)
CVSS Information
N/A
Source: CNNVD (China National Vulnerability Database)
Vulnerability Type
N/A
Source: CNNVD (China National Vulnerability Database)

Affected Products

Vendor Product Affected Versions CPE Subscribe
Medtronic MyCareLink Patient Monitor 24950 0 ~ June 25, 2025 -
Medtronic MyCareLink Patient Monitor 24952 0 ~ June 25, 2025 -

II. Public POCs for CVE-2025-4394

# POC Description Source Link Shenlong Link
AI-Generated POC Premium

No public POC found.

Login to generate AI POC

III. Intelligence Information for CVE-2025-4394

请登录查看更多情报信息。

Vendor Advisories for CVE-2025-4394 (2)

Same Patch Batch · Medtronic · 2025-07-24 · 3 CVEs total

CVE-2025-4395 6.8 MEDIUM Medtronic MyCareLink Patient Monitor Empty Password Vulnerability
CVE-2025-4393 6.5 MEDIUM Medtronic MyCareLink Patient Monitor Deserialization Vulnerability

IV. Related Vulnerabilities

V. Comments for CVE-2025-4394

No comments yet