Goal Reached Thanks to every supporter — we hit 100%!

Goal: 1000 CNY · Raised: 1336 CNY

100%

CVE-2025-14097— Remote Code Execution Vulnerability in Radiometer Products

Quick assessment

Affected
Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers
Exploitation
No confirmed in-the-wild exploitation; assess based on exposure
Recommended action
Check the vendor advisory and references for a fixed version. If immediate upgrade is impossible, restrict exposure and increase monitoring.

Radiometer Products是丹麦Radiometer公司的一系列医疗诊断仪器。 Radiometer Products存在安全漏洞,该漏洞源于分析仪应用软件存在弱点,在满足特定内部条件时可能导致远程代码执行和未经授权的设备管理。

CVSS 7.2 · High EPSS 0.45% · P38
Get alerts for future matching vulnerabilities Log in to subscribe

I. Basic Information for CVE-2025-14097

Vulnerability Information

Have questions about the vulnerability? See if Shenlong's analysis helps!
View Shenlong Deep Dive ↗

Although we use advanced large model technology, its output may still contain inaccurate or outdated information.Shenlong tries to ensure data accuracy, but please verify and judge based on the actual situation.

Vulnerability Title
Remote Code Execution Vulnerability in Radiometer Products
Source: CVE Program / CVE List V5
Vulnerability Description
A vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are met. Exploitation requires that a remote connection is established with additional information obtained through other means. The issue is caused by a weakness in the analyzer’s application software.                                                                                                                                                                                                Other related CVE's are CVE-2025-14095 & CVE-2025-14096.                                                                                                      Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Affected application software version is in use and remote support feature is enabled in the analyzer.                                                                                                                                                                        Temporary work Around: If the network is not considered secure, please remove the analyzer from the network.                         Permanent solution: Customers should ensure the following: • The network is secure, and access follows best practices. Local Radiometer representatives will contact all affected customers to discuss a permanent solution.                                                      Exploit Status: Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any publicly available exploits at the time of this publication.
Source: CVE Program / CVE List V5
CVSS Information
CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H
Source: CVE Program / CVE List V5
Vulnerability Type
认证机制不恰当
Source: CVE Program / CVE List V5
Vulnerability Title
Radiometer Products 安全漏洞
Source: CNNVD (China National Vulnerability Database)
Vulnerability Description
Radiometer Products是丹麦Radiometer公司的一系列医疗诊断仪器。 Radiometer Products存在安全漏洞,该漏洞源于分析仪应用软件存在弱点,在满足特定内部条件时可能导致远程代码执行和未经授权的设备管理。
Source: CNNVD (China National Vulnerability Database)
CVSS Information
N/A
Source: CNNVD (China National Vulnerability Database)
Vulnerability Type
N/A
Source: CNNVD (China National Vulnerability Database)

Affected Products

Vendor Product Affected Versions CPE Subscribe
Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers Application software versions < 3.5MR11 with Windows 7, Windows XP operating systems -
Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX Analyzers Application software versions < 6.20MR2 with Windows 7, Windows XP operating systems -
Radiometer Medical Aps AQT90 FLEX Analyzers Application software versions <= 8.13 MR2 with Windows 7, Windows XP operating systems -

II. Public POCs for CVE-2025-14097

# POC Description Source Link Shenlong Link
AI-Generated POC Premium

No public POC found.

Login to generate AI POC

III. Intelligence Information for CVE-2025-14097

登录查看更多情报信息。

Other References for CVE-2025-14097 (1)

Same Patch Batch · Radiometer Medical Aps · 2025-12-17 · 3 CVEs total

CVE-2025-14096 8.4 HIGH Credential Disclosure vulnerability in Radiometer Products
CVE-2025-14095 5.7 MEDIUM Privilege boundary violation in Radiometer Products

IV. Related Vulnerabilities

V. Comments for CVE-2025-14097

No comments yet


Leave a comment